| Parameter | Wert |
|---|---|
| Artikel | Specifications |
| Gewicht | Net weight: 300 g approx. |
| Details | Controller |
| Produkt | Ultra High-speed Laser Displacement Sensor [CCD Style] |
| Zubehör | - |
| Teilenummer | HL-C1C-M |
| Alarmausgang | Photo-MOS relay- Maximum load current: 50 mA- Applied voltage: 30 V DC or less (between alarm output and COM)- ON impedance: 35 Ohm or less- Operation time: Max. 2 ms |
| Kabellänge | Power line: Less than 10 m 32.808 ftSignal line: Less than 30 m 98.425 ft |
| Produktname | Ultra High-speed Laser Displacement Sensor HL-C1(Discontinued) |
| Abtastrate | Selectable from 100 micro s / 144 micro s / 200 micro s / 255 micro s / 332 micro s / 498 micro s / 1,000 micro s |
| Haltefunktionen | Selectable from NORMAL / P-P / PEAK / VALLEY(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Produktnummer | HL-C1C-M |
| Versorgungsspannung | 24 V DC plus or minus 10 % including ripple 0.5 V (P-P) |
| Digitale Filter | High pass: OFF, 10 to 2,000 Hz (9 steps)Low pass: OFF, 10 to 2,000 Hz (9 steps)(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Indikatoren: DUNKEL | Red LED (lights up upon disabled measurement due to insufficient light at sensor head 1 or 2) |
| Nullpunkt gesetzt EIN Eingang | Zero set: ON when voltage (using input voltage: 12 to 24 V DC, maximum input voltage: 30 V DC) is input |
| Kalibrierung: Spanne | 0.9000 to 1.1000(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Nullpunkt-OFF-Eingang | Zero set: OFF when voltage (using input voltage: 12 to 24 V DC, maximum input voltage: 30 V DC) is input |
| Kalibrierung: Verschiebung | plus or minus 20.0000 mm plus or minus 0.787 in(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Aktueller Verbrauch | When 1 sensor is connected: 430 mA approx.When 2 sensors are connected: 550 mA approx. |
| Indikatoren: HELL | Red LED (lights up upon disabled measurement due to excessive light at sensor head 1 or 2) |
| Berechnungsfunktionen | L plus or minus KA, L plus or minus KB, L plus or minus K (A plus or minus B)A, B: Sensor head 1, Sensor head 2 measurement values, L = plus or minus 999.9999, K = 0.0001 to 99.9999(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Serielle Ein-/Ausgabe | RS-232C |
| Connectale Sensorkopf | Maximum 2 sensor heads |
| Analogausgang: Strom | Output current: 4 to 20 mA/F.S. [default setting when diffuse reflective mode is selected (Note 2)]Output range: 0 to 29.5 mA (maximum of 25 mA at max. load impedance)Load impedance: 250 Ohm or less(Note 1) The maximum analog output current will vary with load impedance.(Note 2) If specular reflective mode is selected, then the default setting is 5.6 to 18.4 mA/F.S. |
| Analogausgang: Spannung | Output voltage: plus or minus 5 V/F.S. [default setting when diffuse reflective mode is selected (Note) ]Output range: -10.9 to +10.9 VOutput current: Max. 2 mA, Output impedance: 50 Ohm(Note) If specular reflective mode is selected, then the default setting is plus or minus 4 V/F.S. |
| Durchschnittliche Anzahl der Proben | OFF, 2 to 32,768 times (16 steps)(Note) These values can be set using the command input from external equipment via the compact console and RS-232C interface. |
| Beurteilungsergebnisse (O1, O2) | Photo-MOS relay- Maximum load current: 50 mA- Applied voltage: 30 V DC or less (between judgment output and COM)- ON impedance: 35 Ohm or less- Operation time: Max. 2 ms |
| Indikatoren: Laseremission | Green LED (lights up during laser emission from sensor head 1 or sensor head 2, or immediately before laser emission) |
| Temperatureigenschaften | plus or minus 0.01 % F.S./degrees |
| Zeiteingang (Laseremission) | Laser emission stops or continues when voltage (using input voltage: 12 to 24 V DC, maximum input voltage: 30 V DC) is input or there is an open circuit: determined based on input mode selection. |
| Alarmausgang: Ausgangsfunktion | Opened when the amount of light is excessive or insufficient. |
| Alarmausgang: Kurzschlussschutz | Incorporated |
| Umweltbeständigkeit: Umgebungsfeuchtigkeit | 35 to 85 % RH, Storage: 35 to 85 % RH |
| Umweltbeständigkeit: Stoßfestigkeit | 196 m/s2 (20 G approx.) in X, Y and Z directions for 3 times each |
| Beurteilungsergebnisse (O1, O2): Ausgabeoperation | Opened or closed when the threshold value is reached. Determined based on judgment output mode selection. (The threshold value varies with the hysteresis setting.) |
| Umweltbeständigkeit: Umgebungstemperatur | 0 to +50 degrees +32 to +122 degrees Fahrenheit (No dew condensation), Storage: -20 to +70 degrees -4 to +158 degrees Fahrenheit |
| Umweltbeständigkeit: Vibrationsbeständigkeit | 10 to 55 Hz frequency (period: 1 min.) 0.75 mm 0.030 in amplitude in X,Y and Z directions for 30 min. each |
| Beurteilungsergebnisse (O1, O2): Nutzungskategorie | DC-12 or DC-13 |
| Ausgänge für die Beurteilung (O1, O2): Kurzschlussschutz | Incorporated |
Technische Spezifikationen und Leistungsdaten