| Parameter | Wert |
|---|---|
| Artikel | Specifications |
| Gewicht | Net weight: 155 g approx., Gross weight: 220 g approx. |
| Details | Aluminum body type |
| Anzeige | 4 digits + 4 digits 2-color LCD display (Display refresh rate: 250 ms, 500 ms, 1,000 ms, selectable by key operation) |
| Produkt | Integrated Display Type Digital Flow Sensor |
| Material | Enclosure: ABS, Body: Polyamide (Aluminum body type: Aluminum), Switch: EPDM, Display: Acrylic, Mounting screw part (Resin body type): Brass Current plate / Port filter: Stainless steel (used for the gas contact area), Sensor chip: Silicon, Gasket: Fluorine rubber |
| Linearität | Within ±3 % F.S. (Ambient temperature +25 ℃ +77 ℉, flow rate range 3 to 100 % F.S., atmospheric criteria on secondary side) |
| Portgröße | Rc1/2 female thread |
| Schutz | IP40 (IEC) |
| Zubehör | CN-F15-C1 (Connector attached cable 1 m 3.281 ft): 1 pc. |
| Teilenummer | FM-216-AR2 |
| Kabellänge | Total length up to 10 m 32.808 ft is possible with 0.3 mm2, or more, cable. |
| Produktname | Integrated Display Type Digital Flow Sensor FM-200 |
| Wiederholbarkeit | Within ±1 % F.S. |
| Externer Eingang | ON voltage: 0 to +0.4 VOFF voltage: +5 V to +V, or openInput time: 80 ms or more |
| Produktnummer | FM-216-AR2 |
| Versorgungsspannung | 12 to 24 V DC ± 10 % Ripple P-P10 % or less |
| Umgebungsfeuchtigkeit | 35 to 90 % RH, Storage: 35 to 90 % RH |
| Geeignete Flüssigkeit | Clean air, compressed air, nitrogen gas(Note):The clean air complies with JIS B 8392-1.1.1 to 5.6.2, and the compressed air complies with JIS B 8392-1.1.1 to 1.6.2. |
| Stoßfestigkeit | 100 m/s2acceleration (10 G approx.) in X, Y and Z directions three times each |
| Verbindungsmethode | Connector |
| Vergleichsausgabe | NPN open-collector transistor・Maximum sink current: 50 mA or less・Applied voltage: 26.4 V DC or less (between comparative output and 0 V)・Residual voltage: 2.4 V or less (at 50 mA sink current) |
| Gehäuseerdung | Floating(Note):As a varistor (clamping voltage: approx. 40 V) is connected to the aluminum body type, do not apply voltage higher than the rated voltage of the sensor. |
| Umgebungstemperatur | 0 to +50 ℃ +32 to +122 ℉ (No dew condensation allowed), Storage: -10 to +60 ℃ +14 to +140 ℉ |
| Aktueller Verbrauch | Normal mode: 60 mA or less, ECO mode: 40 mA or less |
| Durchflussrate im Vollausschlag | 1,000 ℓ/min.(Note):Converted to volumetric flow at +20 ℃ +68 ℉ and 1 atmospheric pressure (101 kPa). |
| Nenndruckbereich | -0.09 to +0.7 MPa |
| Vibrationsfestigkeit | 10 to 150 Hz frequency, 0.75 mm0.030 indouble amplitude or 49 m/s2max. acceleration, in X, Y and Z directions for two hours each |
| Analoger Spannungsausgang | Output voltage: 1 to 5 V, Output impedance: 1 kΩ approx.[Refer to "Analog voltage output" for more details.] |
| Isolationswiderstand | 10 MΩ, or more, with 500 V DC megger between all supply terminals connected together and enclosure (Excluding the aluminum body type) |
| Genauigkeitssicherungsbereich | Bi-direction : -100 to -3 % F.S., +3 to +100 % F.S.One-side direction : +3 to +100 % F.S.(Note):Take care that if fluid flows in the vicinity of zero-point which is out of the accuracy assurance range, the instant flow rate value may forcibly display "zero", or the integrated flow display value may not be counted up, or the integrated pulse output may not be outputted. |
| Druckcharakteristik | Within ±5 % F.S. (-0.09 to +0.7 MPa, +25 ℃ +77 ℉, atmospheric criteria on secondary side) |
| Spannungsfestigkeit | 1,000 V AC for one min. between all supply terminals connected together and enclosure (Excluding the aluminum body type) |
| Druckbeständigkeit | 1 MPa |
| Angegebener integrierter Wert | 10ℓ |
| Temperaturcharakteristik | Within ±0.2 % F.S./℃ (+25 ℃ +77 ℉ criteria, +15 to +35 ℃ +59 to +95 ℉ ambient temperature range) |
| Vergleichsausgabe: Hysterese | Window comparator mode: 1 to 8 % F.S. approx. (variable) (Factory settings: approx. 1 % F.S.), Hysteresis mode: Variable (minimum 1 digit) |
| Einhaltung der CE-Kennzeichnungsrichtlinie | EMC Directive, RoHS Directive |
| Vergleich der Ausgänge: Ausgabemodi | Output OFF mode, window comparator mode, hysteresis mode, integrated output mode, integrated pulse output mode |
| Momentaner Durchfluss: Anzeigebereich | -1,100 to +1,100 ℓ/min.(Note):The display flow rate range is the case when setting to bi-direction at the flow direction setting. When the flow direction is set to one-side forward direction or one-side reverse direction, the negative side of the display flow rate range shows 10 % of the full-scale (F.S.). |
| Vergleichsausgabe: Reaktionszeit | 50 ms, 80 ms, 120 ms, 200 ms, 400 ms, 800 ms, 1,500 ms, selectable by key operation |
| Integrierter Durchfluss: Anzeigebereich | ±9999999 ℓ(Note):The display flow rate range is the case when setting to bi-direction at the flow direction setting. When the flow direction is set to one-side forward direction or one-side reverse direction, the negative side of the display flow rate range shows 10 % of the full-scale (F.S.). |
| Vergleichsausgang: Kurzschlussschutz | Incorporated |
| Momentaner Durchfluss: Einstellung und Anzeigeauflösung | 1 ℓ/min. |
| Integrierter Durchfluss: Einstellung und Anzeigeauflösung | 1 ℓ |
Technische Spezifikationen und Leistungsdaten
Installations- und Betriebsanleitung